Abstract

The aim of the study was to examine the clinical outcome of fixing unstable intertrochanteric fractures using a newly designed 102° dynamic condylar screw (DCS). Twelve unstable low-energy intertrochanteric fractures were fixed using 102° DCS. Patient ages ranged from 79 to 92 years. The follow-up period was from 6 to 15 months. Patients were assessed clinically and radiographically. Fracture healing was achieved in 11 patients within the follow-up period. Implant failure was seen in one patient, as the screw cut out of the femoral head. In conclusion, the 102° DCS showed promising results when used for fixing unstable intertrochanteric fractures. It is easier to apply in comparison to the 95° DCS, as it offers wider safety zone for insertion in the femoral neck, which allows a secure screw placement in the femoral head. Larger randomised comparative studies would be of value for further evaluation of the new device.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.