Abstract

During storage, the potassium level of red blood cell (RBC) components increases, especially after irradiation. Neonates are prone to hyperkalemia, for example, non-oliguric hyperkalemia, so using potassium adsorption filters during transfusion may be helpful. To overcome dilution of RBC components caused by saline priming of existing potassium adsorption filters, a downsized potassium adsorption filter for neonates (PAF-n, Kawasumi Laboratories Inc., Tokyo, Japan) was developed. To assess the performance of PAF-n, its adsorption efficiency and RBC recovery rate were evaluated by testing pre-filtration and serial post-filtration (0-30 mL, 30-60 mL, 60-90 mL, and 90-120 mL) samples from 8 RBC components. The average potassium adsorption rate of the PAF-n was 90.5% ± 0.78%, and never less than 89.0% in any of 8 RBC components. RBC recovery rates were 99.3% ± 1.12%. The PAF-n showed an effective potassium ability with negligible RBC dilution.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.