Abstract

The improvement in detection limits for low and high activity samples measured with the Compton-suppression γ-spectrometer installed at the Institute for Transuranium Elements (ITU) for environmental monitoring of radioactivity, as well as nuclear safeguards, is discussed. The advantage of using two parallel acquisition lines for simultaneous measurement with and without Compton-event suppression is outlined with respect to cascade and non-cascade γ-emitters. The background reduction by Compton-event suppression made it possible to detect small peaks, which otherwise would not have been found in a conventional spectrum. In Compton-event suppression mode, the detection limit for 137Cs was improved by a factor of about 3, for 241Am we found a factor of 1.2 both in high and low active samples. The measurements of environmental reference samples showed good agreement with certified values in both acquisition modes. The application of this instrument for the determination of fission products in smear samples is described. In particular, for nuclides like 54Mn, 125Sb and 144Ce, an improvement in the detection limits by a factor of 1.6–2.4 was obtained.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.