Abstract

AbstractA method is introduced to assess and correct the geometric distortions which frequently occur in low‐magnification scanning electron microscopy (SEM) images. Such images typically exhibit a complex pattern of varying deviations from orthogonality which cannot be adequately corrected by simple geometric transformations such as shifting, scaling, rotation, or shearing. A suitable approach to rectify low‐magnification SEM images is polynomial warping, a correction procedure which also accomplishes rubber sheet transformation. To demonstrate the approach, a reference grid for low magnifications has been scanned at 40‐ and 55‐fold magnifications by means of a microanalyzer. Calculated geometric distortions range from 1.5 to 3.5% of the image dimensions; applying polynomial warping, distortions could be reduced to approximately 0.1% of the image dimensions. Because of its easy application and the widespread availability in image processing packages, polynomial warping can be recommended as a routine procedure for rectifying low‐magnification SEM images.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.