Abstract

This study compared the synchronisation of a medical grade Electroencephalography (EEG) system, the g.Tec, and a consumer grade EEG system, the Emotiv. Data was collected from both systems using the lab streaming layer (LSL). Both EEG systems recorded an electric signal from the surface of a customised gel phantom. The electric signal was generated using a solar cell which was illuminated by a monitor presenting a sequence of black and white images. Test results show that the g.Tec had a mean delay of 51.22 ms from the stimulus onset and the Emotiv had a mean delay of 162.69 ms from the stimulus onset. The result should be taken into account with future ERP studies which will use either the EEG system and the lab streaming layer. The design of this experiment provides a smart way to evaluate the temporal accuracy of other EEG systems.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.