Abstract

Background: Quality control (QC) of computed tomography (CT) scanners is important to evaluate succinctly quality image and radiation dose obtainable in a clinical environment. The aim of this study was to evaluate the quality of images generated by CT scanners used at some diagnostic facilities in Ibadan, Nigeria.
 Materials and Methods: A cross sectional design was employed in this study, four centers were studied, one government hospital and three private hospitals. The head CT phantom was used to verify the accomplishment of the CT scanners performance to the international quality requirements. Regions of interest were selected at the center of the image and at the periphery to obtain results for the CT number for water test, uniformity test, noise, and artifact test.
 Results: The mean CT number for water across the centers ranged from –0.12 HU to –2.2 HU which were within ±3 HU recommended by the equipment manufacturer. Values of standard deviation of the mean CT number ranged from 2.41 to 5.77 HU which to a little extent exceeded the set ±5 HU tolerance range. Similarly, the presence of streak artifact was observed in the images obtained at one center.
 Conclusion: Two out of the four computed tomography scanners assessed passed the four tests performed. Noise and artifact were the problem observed at centers B and C respectively. There was however no likelihood of periodic performance of these basic quality control tests at two of the centers in this study. Adequate records of quality control data should be kept regularly to allow in-depth analysis of failure rates of different tests, changes occurring during equipment lifetime and comparisons among CT scanners.

Highlights

  • Computed tomography (CT) is a medical x-ray imaging modality that transmits numerous x-ray beams through the human body to obtain images for disease diagnosis, evaluation of clinical treatment, and assessment of patient wellness [1]

  • The mean computed tomography (CT) number for water across the centers ranged from –0.12 HU to –2.2 HU which were within ±3 HU recommended by the equipment manufacturer

  • Adequate records of quality control data should be kept regularly to allow in-depth analysis of failure rates of different tests, changes occurring during equipment lifetime and comparisons among CT scanners

Read more

Summary

Introduction

Computed tomography (CT) is a medical x-ray imaging modality that transmits numerous x-ray beams through the human body to obtain images for disease diagnosis, evaluation of clinical treatment, and assessment of patient wellness [1]. This is mainly significant for the diagnosis of pathological conditions involving soft tissue. One important element of taking care of equipment is to conduct Quality Control tests. Quality control (QC) of computed tomography (CT) scanners is important to evaluate succinctly quality image and radiation dose obtainable in a clinical environment. The aim of this study was to evaluate the quality of images generated by CT scanners used at some diagnostic facilities in Ibadan, Nigeria

Objectives
Results
Discussion
Conclusion
Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call