Abstract
In this work, we report installation of displex cryostat XAFS sample holder at XAFS beamline (BL-09) of Indus-2 synchrotron facility, India and make critical assessment of feasibility of low-temperature XAFS experiments in terms of data quality and reproducibility, temperature range, calibration and attainable resolution. We adopted the Debye Model-based calibration method by measuring XAFS of standard Au foil with known Debye temperature (ΘDebye)Autheory=165K. The data is of good quality and reproducible with international data. By fitting Debye Waller Factor (σexpt2(T)), we deduced (ΘDebye)Auexpt=163K which implies calibration within 2K. Error bars for σexpt2(T) correspond to temperature uncertainty ΔT≤5K, which defines the temperature resolution for low temperature XAFS experiments. Thus, from both calibration and resolution points-of-view, this work demonstrates the feasibility of low temperature XAFS experiments at BL-09, Indus-2. Feasibility of extending XAFS experiments to lower temperature and unknown samples is discussed.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.