Abstract
The design and manufacture of intraocular lenses (IOLs) depend upon the identification and quantitative preclinical evaluation of key optical properties and environmental parameters. The confocal laser method (CLM) is a new technique for measuring IOL optical properties, such as dioptric power, optical quality, refractive index, and geometrical parameters. In comparison to competing systems, the CLM utilizes a fiber-optic confocal laser design that significantly improves the resolution, accuracy, and repeatability of optical measurements. Here, we investigate the impact of changing the beam diameter on the CLM platform for the evaluation of IOL dioptric powers. Due to the Gaussian intensity profile of the CLM laser beam, the changes in focal length and dioptric power associated with changes in beam diameter are well within the tolerances specified in the ISO IOL standard. These results demonstrate some of the advanced potentials of the CLM toward more effectively and quantitatively evaluating IOL optical properties.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.