Abstract

BackgroundWe developed SmartRub® powered by iQati®, an electronic device composed of a wristband and an alcohol-based handrub pocket-sized dispenser that measures and provides feedback on the duration of hand friction and the volume poured during each hand hygiene action. We aimed to assess the accuracy of SmartRub®.MethodsThe specificity, sensitivity, positive and negative predictive values (PPV and NPV) of SmartRub® were assessed in a 3-phased experiment: (1) laboratory-controlled conditions with volunteers; (2) pre-planned clinical path with volunteers and (3) real clinical conditions with healthcare workers. The accuracy of SmartRub® was evaluated by quantifying its ability to correctly capture true hand hygiene actions and to not record other actions performed while wearing the device.ResultsIn the laboratory, 7 volunteers performed 816 actions. Overall sensitivity was 94.1% (95% CI 91.4–96.2%) with a PPV of 99.0% (95% CI 97.3–99.6%) and specificity was 99.0% (95% CI 97.5–99.7%) with a NPV of 94.4% (95% CI 91.9–96.1%). During the pre-planned clinical path phase, 13 volunteers performed 98 planned paths and a total of 967 actions were performed. Overall sensitivity was 94.6% (95% CI 92.2–96.5%) with a PPV of 84.3% (95% CI 81.6–86.7%) and specificity was 82.4% (95% CI 78.7–85.7%) with a NPV of 93.9% (95% CI 91.3–95.7%). During the real clinical conditions phase, 17 healthcare workers were observed for a total of 15 h and 3 min while they performed 485 actions. Sensitivity was 96.8% (95% CI 93.8–98.6%) with a PPV of 98.3% (95% CI 95.6–99.3%) and specificity was 98.3% (95% CI 95.7–99.5%) with a NPV of 96.8% (95% CI 93.9–98.4%).ConclusionsSmartrub® is a highly reliable device for capturing hand hygiene actions under a range of conditions, from the laboratory to clinical care activities.

Highlights

  • We developed SmartRub® powered by iQati®, an electronic device composed of a wristband and an alcohol-based handrub pocket-sized dispenser that measures and provides feedback on the duration of hand friction and the volume poured during each hand hygiene action

  • Overall sensitivity was 94.1% with a PPV of 99.0% and specificity was 99.0% with a NPV of 94.4%

  • Overall sensitivity was 94.6% with a PPV of 84.3% and specificity was 82.4% with a NPV of 93.9%

Read more

Summary

Introduction

We developed SmartRub® powered by iQati®, an electronic device composed of a wristband and an alcohol-based handrub pocket-sized dispenser that measures and provides feedback on the duration of hand friction and the volume poured during each hand hygiene action. To complement the WHO method of direct observation, some healthcare facilities are using additional interventions such as electronic dispensing counters and automated HH monitoring networks (including wearable devices) [7]. These automated systems likely require fewer human resources, provide larger and more representative data sets, and are less subject to observation bias than the WHO recommended direct observation method [8,9,10]. Healthcare workers (HCWs) frequently identify lack of system accuracy as undermining their trust in the tool [11]

Objectives
Methods
Results
Conclusion