Abstract

BackgroundPreventive interventions save lives during the process of chemotherapy for hematologic malignancies, when a hematology laboratory can ensure accurate results. The use of a pneumatic tube system (PTS) is associated with measurement errors and unnecessary transfusions. The aim of this study was to evaluate pre-analytical errors associated with transportation method (PTS versus hand-delivered) and to investigate whether there are unnecessary transfusion events in pancytopenia leukemia patients with very low hematological parameters.Material/MethodsA total of 140 paired blood collections were performed for hemogram and biochemistry assays. Paired EDTA and serum gel blood samples were collected from 58 cases with acute leukemia on different days. For each pair, one sample was hand-delivered by a courier (Group 1) while the other sample was transported through a PTS (Group 2).ResultsThe hand-delivered method showed that some platelet transfusions were unnecessary for different thrombocyte cut-off values. Calculated unnecessary platelet (PLT) transfusion ratios when using PTS (PLT <30×103/μL, 16.3%; PLT <25×103/μL, 16.4%; PLT <20×103/μL, 80.3%; PLT <15×103/μL, 48.6%; and PLT <10×103/μL, 150.0%) were found to be statistically significant (p=0.002, p=0.046, p<0.000, p=0.028, and p<0.000, respectively). In contrast, for RBC transfusion ratios, although the ratios were high in Group 2, we found no significant difference between the two groups; (HGB <8.0 g/dL, 23.3%; HGB <9.0 g/dL, 25.0%, HGB<10.0 g/dL, 19.3%) and (p=0.002, p=0.085, p<0.160, and p=0.235, respectively).ConclusionsAlthough our results cannot be universally applied, physicians should be careful, skeptical, and suspicious of transfusion decisions in hematology clinics and consider potential analytical and pre-analytical errors in cases of severe cytopenia when using PTS.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call