Abstract

Photoplethysmographic imaging (PPGI) systems are relatively new non-contact biophotonic diffuse reflectance systems able to assess arterial pulsations through transient changes in light-tissue interaction. Many PPGI studies have focused on extracting heart rate from the face or hand. Though PPGI systems can be used for widefield imaging of any anatomical area, whole-body investigations are lacking. Here, using a novel PPGI system, coded hemodynamic imaging (CHI), we explored and analyzed the pulsatility at major arterial locations across the whole body, including the neck (carotid artery), arm/wrist (brachial, radial and ulnar arteries), and leg/feet (popliteal and tibial arteries). CHI was positioned 1.5 m from the participant, and diffuse reactance from a broadband tungsten-halogen illumination was filtered using 850{1000 nm bandpass filter for deep tissue penetration. Images were acquired over a highly varying 24-participant sample (11/13 female/male, age 28.7±12.4 years, BMI 25.5±5.2 kg/m 2 ), and a preliminary case study was performed. B-mode ultrasound images were acquired to validate observations through planar arterial characteristics.