Abstract

BackgroundWith the development and promotion of wearable devices and their mobile health (mHealth) apps, physiological signals have become a research hotspot. However, noise is complex in signals obtained from daily lives, making it difficult to analyze the signals automatically and resulting in a high false alarm rate. At present, screening out the high-quality segments of the signals from huge-volume data with few labels remains a problem. Signal quality assessment (SQA) is essential and is able to advance the valuable information mining of signals.ObjectiveThe aims of this study were to design an SQA algorithm based on the unsupervised isolation forest model to classify the signal quality into 3 grades: good, acceptable, and unacceptable; validate the algorithm on labeled data sets; and apply the algorithm on real-world data to evaluate its efficacy.MethodsData used in this study were collected by a wearable device (SensEcho) from healthy individuals and patients. The observation windows for electrocardiogram (ECG) and respiratory signals were 10 and 30 seconds, respectively. In the experimental procedure, the unlabeled training set was used to train the models. The validation and test sets were labeled according to preset criteria and used to evaluate the classification performance quantitatively. The validation set consisted of 3460 and 2086 windows of ECG and respiratory signals, respectively, whereas the test set was made up of 4686 and 3341 windows of signals, respectively. The algorithm was also compared with self-organizing maps (SOMs) and 4 classic supervised models (logistic regression, random forest, support vector machine, and extreme gradient boosting). One case validation was illustrated to show the application effect. The algorithm was then applied to 1144 cases of ECG signals collected from patients and the detected arrhythmia false alarms were calculated.ResultsThe quantitative results showed that the ECG SQA model achieved 94.97% and 95.58% accuracy on the validation and test sets, respectively, whereas the respiratory SQA model achieved 81.06% and 86.20% accuracy on the validation and test sets, respectively. The algorithm was superior to SOM and achieved moderate performance when compared with the supervised models. The example case showed that the algorithm was able to correctly classify the signal quality even when there were complex pathological changes in the signals. The algorithm application results indicated that some specific types of arrhythmia false alarms such as tachycardia, atrial premature beat, and ventricular premature beat could be significantly reduced with the help of the algorithm.ConclusionsThis study verified the feasibility of applying the anomaly detection unsupervised model to SQA. The application scenarios include reducing the false alarm rate of the device and selecting signal segments that can be used for further research.

Highlights

  • BackgroundWearable devices have been widely adopted for daily health care monitoring during the past decades

  • The algorithm application results indicated that some specific types of arrhythmia false alarms such as tachycardia, atrial premature beat, and ventricular premature beat could be significantly reduced with the help of the algorithm

  • This study verified the feasibility of applying the anomaly detection unsupervised model to Signal quality assessment (SQA)

Read more

Summary

Introduction

BackgroundWearable devices have been widely adopted for daily health care monitoring during the past decades. Some clinical deterioration changes in physiological signals could be often present 8-24 hours before a severe life-threatening event such as an unplanned intensive care unit admission or sudden cardiac death [8,9]. In these scenarios, signal quality is essential to acquire the valuable information from the time-series physiological signals which are very sensitive to noise. Signal quality assessment (SQA) facilitates reducing the high false alarm rate caused by signal quality [10] and can be applied to automatically screen the “real-world” data for further research. Signal quality assessment (SQA) is essential and is able to advance the valuable information mining of signals

Methods
Results
Discussion
Conclusion