Abstract

AbstractFine PTFE tubes are used in high‐end and high cost devices like flexible Endoscopes where no standard sterilization procedure like autoclaving applies due to material properties. The study of innovative decontamination processes like plasma sterilization request appropriate analysis methods in order to verify the process functionality. In a related research project the IGB‐cycle‐method for contamination and recovery of indicator organisms like Bacillus atrophaeus spores was developed. It was possible to proof the IGB‐cycle‐method to be reproducible and reliable using round robin tests between three laboratories that compared their results in reference to a relevant level of contamination and recovery. Of a relevant contamination more than 7 log10 steps of cfu/specimen could be recovered from the tube surface. magnified image

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.