Abstract
We evaluated three commercially available methods for determining lipase (EC 3.1.1.3) in serum--the Du Pont aca, Boehringer Mannheim Diagnostics (BMD), and Kodak Ektachem (EK) procedures--for their analytical properties and diagnostic efficiencies. Titrimetry was used as the comparative method. The BMD and EK methods showed better agreement with the titrimetric method, owing to the presence of the necessary cofactor, colipase, in their reagents. Colipase also increased the analytical sensitivity of the BMD and EK procedures as compared with the aca method. Determinations of serum lipase, by all methods, had a clinical sensitivity in excess of 80% for acute pancreatitis; the specificity of the lipase test was about 60%, or twice that of serum amylase. Serum lipase determinations with the current, simpler technology are superior to total amylase in the diagnosis of patients with acute pancreatitis. When a colipase-supplemented method is used, a serum lipase value greater than 10-fold the upper reference limit appears to be pathognomonic for acute pancreatitis or inflammation of organs close to the pancreas.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.