Abstract

For the precise measurements of special nuclear materials (SNM) including Pu and Am isotopes, we have used phoswich detector combination of two single crystal scintillators of Gd3Ga3Al2O12:Ce and CsI:Tl. High detection efficiency and sensitivity along with high figure of merit for the discrimination of these phoswich detectors ensures the detection and discrimination of thermal neutrons and gammas from spontaneous fission of Pu and other isotopes in presence of high gamma background. Using this detector, the low energy gammas, which is stopped completely in 1mm thick disc of GGAG, can be also discriminated from high energies gamma and shows linearity in wide range of sample quantities. By changing only the appropriate shielding, the similar setup was used for thermal neutron detection and shows a very good linearity over wide range. The quantity of a test sample was also calculated accurately by using the measured calibrated plot.