Abstract

A technique is presented for measuring the 2H enrichment of water in biological samples when this enrichment is greater than 0.2%. The sample is reacted with calcium carbide to form acetylene gas, which is determined by gas chromatography electron impact ionization mass spectrometry. Ion-molecule reactions, resulting in proton abstraction, are minimized by lowering the electron ionization energy from the usual 70 eV to 45 eV. This technique is much more rapid and economical than the classical isotope ratio mass spectrometric assay of the enrichment of hydrogen gas derived from reduction of water.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.