Abstract

A novel in-vitro system has been developed for the detection and quantification of pyrogen in pharmaceutical products. The measured variable was evoked secretion of the pyrogenic cytokine interleukin-6 from MONO MAC 6 monocytic cells incubated with the product. The interleukin-6 was detected using a specific and sensitive ELISA developed for this purpose. The test system detected pyrogenic contamination in 3 batches of therapeutic human serum albumin which had caused adverse reactions in recipients. The contamination was not detected in conventional tests: the rabbit pyrogen test and the limulus amoebocyte lysate test.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.