Abstract

In the present study, we tried to reduce the non-specific reactions for measuring anti-Chlamydia pneumoniae IgM antibodies by the ELISA kit of HITAZYME C. pneumoniae Ab-IgM (HITAZYME IgM) by using a new absorbent. We also tried to reset the IgM cut-off index (ID) of HITAZYME IgM by testing serum samples from healthy children and healthy adults with no respiratory symptoms. The results suggest that the use of the new absorbent (anti-human IgG antibodies) may reduce the non-specific reactions by rheumatoid factor and anti-nuclear antibodies, and that the setting of the higher cut-off ID (2.00), calculated as the mean ID + 3SD of the serum samples from healthy children and healthy adults, respectively, would improve the specificity of IgM during the measurement by HITAZYME IgM.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.