Abstract

A simplified cross-calibration procedure used in laser dynamic goniometry is examined, in which the angular displacement increment of the studied angle sensor exceeds its resolution. The obtained method error for M uniform angular increments of the angle sensor is determined by the overall amplitude of the Mth harmonic and its aliquot harmonics of the goniometer systematic error. It is demonstrated experimentally that when the angular displacement increment of an angle sensor is less than 120°, the method error does not exceed 0.01′′.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.