Abstract
A sample received in the laboratory from a patient receiving total parenteral nutrition (TPN) indicated that the patient may have renal dysfunction, but the results were not considered to be reliable enough to report. Investigations using a reference method for measurement of creatinine confirmed positive interference in the creatinine assay and distribution of samples via an External Quality Assessment (EQA) Scheme showed that this positive interference was method dependent. Residual TPN fluid (Nutriflex Lipid Special) left in the bag after the patient had completed the infusion was collected and added to a patient serum pool in increasing amounts and distributed to different laboratories for analysis of creatinine and glucose through an EQA Scheme. Positive interference in a number of different creatinine assays was identified as a result of a component in the TPN fluid. Positive interference from high concentrations of glucose has been demonstrated to be a cause for falsely high results in Jaffe creatinine assays. The concern would be that a sample contaminated with TPN fluid would have both abnormal electrolytes and creatinine concentrations and give the impression that the patient was in renal failure due to analytical interference in the creatinine assay and laboratory staff need to be aware of this problem.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Annals of Clinical Biochemistry: International Journal of Laboratory Medicine
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.