Abstract
An artificial liver support system for plasma exchange and plasma perfusion through BR-601 resin using a membrane separator was applied to 5 patients with postoperative liver failure. Percent absorption of total and direct bilirubin, and of bile acids were 77.1 +/- 6.4, 78.4 +/- 6.1, and 93.4 +/- 3.6%, respectively, when 250 ml of plasma was treated. Percent reductions in total and direct bilirubin, and in bile acids were 24.5 +/- 5.8, 25.5 +/- 5.8 and 30.9 +/- 8.5%, respectively. In contrast, percent reductions in total and direct bilirubin, and in bile acids by plasma exchange were 30.9 +/- 13.3, 34.5 +/- 12.5 and 24.2 +/- 8.5%, respectively. The coma grade was improved in 4 out of 5 cases, but unfortunately the patients did not recover. In conclusion, plasma perfusion through BR-601 resin is expected to play a promising role in artificial liver support systems because of its capacity to absorb bilirubin and bile acids.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.