Abstract

PurposeFor patients with neuromuscular disease, air stacking, which inflates the lungs to deep volumes, is important for many reasons. However, neuromuscular patients with severe glottic dysfunction or indwelling tracheostomy tubes cannot air stack effectively. For these patients, we developed a device that permits deep lung insufflations substituting for glottic function.Materials and MethodsThirty-seven patients with bulbar-innervated muscle weakness and/or tracheostomies were recruited. Twenty-three had amyotrophic lateral sclerosis, and 14 were tetraplegic patients due to cervical spinal cord injury. An artificial external glottic device (AEGD) was used to permit passive deep lung insufflation. In order to confirm the utility of AEGD, vital capacity, maximum insufflation capacity (MIC), and lung insufflation capacity (LIC) with AEGD (LICA) were measured.ResultsFor 30 patients, MICs were initially zero. However, with the use of the AEGD, LICA was measurable for all patients. The mean LICA was 1,622.7±526.8 mL. Although MIC was measurable for the remaining 7 patients without utilizing the AEGD, it was significantly less than LICA, which was 1,084.3±259.9 mL and 1,862.9±248 mL, respectively (p<0.05).ConclusionThe AEGD permits lung insufflation by providing deeper lung volumes than possible by air stacking.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call