Abstract

Artificial disc replacement was performed in six patients with an average age of 55 years and average follow-up of 3.4 years. Four of the six patients had juxtafusion degeneration, one had multilevel disc degeneration, and one patient had isolated disc resorption. The Acroflex disc (Acromed Corporation, Cleveland, OH), which was used in the replacement, is composed of a rubber core vulcanized to two titanium end plates. The latter have superior posts to provide for initial mechanical fixation and porous in-growth surfaces for long-term fixation. Satisfactory results occurred in four of six patients. Poor results occurred in the presence of deformity that resulted in prosthetic failure and isolated disc resorption. There was an average 8 degrees angular and 2.3-mm translational movement and satisfactory in-growth at all interfaces. Design objectives of endurance, biocompatibility, geometry, kinematics, constraint, dynamics, stability, and fail-safe were met; however, this study is preliminary in nature.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.