Abstract

During computed tomography (CT) pelvimetry of two pregnant women, bony abnormalities of the fetuses were noted on the scout images that were not confirmed at delivery. To explore the cause of these artifacts, specimen long bones were manipulated in various ways during CT scout imaging. Artifacts like those seen during in vivo imaging were found to be caused by motion of the object. The CT scout view is an example of an image produced by a digital system that uses a scanning beam. This type of digital system is being used for several types of body imaging including screening for scoliosis and chest radiography. Attention to motion artifacts should decrease diagnostic errors and aid further development of these systems.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.