Abstract
Unexpectedly large fluctuations in signal intensity were identified in the functional MRI (FMRI) of normal subjects breathing pure oxygen intermittently. To test the hypothesis that the signal changes were due to fluctuating concentrations of gaseous (paramagnetic) oxygen in the magnetic field, echo planar gradient echo images were acquired of a phantom contiguous to an oxygen mask through which pure oxygen was administered intermittently via plastic tubing. As a control, room air was administered intermittently or oxygen continuously in the same experimental protocol. Signal intensity changes of up to 60% temporally correlated with the administration of oxygen were produced in the phantom. In functional images prepared from the echo planar images, the signal intensity changes resulted in artifacts especially at interfaces in the phantom. The intermittent administration of pure oxygen during acquisition of data for FMRI may produce signal intensity changes that stimulate or obscure function.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.