Abstract

Auger electron spectroscopy analyses of submicron features on semiconductor surfaces are routinely accompanied by analytical artifacts such as sample degradation and background contributions arising from electron beam scattering. Submicron analyses are commonly carried out at electron beam densities in excess of 1 A cm -2 and are especially damaging to silicon oxides. The evolution of oxide reduction is observed both as a loss of oxygen versus beam exposure and in a complementary growth of Si LVV and Si KLL elemental peaks. The O KLL signal intensity from a 1 μm 2 area of thermally grown oxide is found to decrease by 22% after exposure to a rastered 20 kV/10 nA beam for 10 min. Another aspect of submicron analysis is the contribution to survey spectra that originates when surrounding material is excited by backscattered electrons. Background contributions may dominate AES spectra even when the sample is flat and the probing beam is smaller than the feature of interest. Tungsten damascene contacts provide a useful platform for investigating this phenomenon in the absence of topography. Spectra have been collected from tungsten contacts of various sizes and the target and background contributions quantified. When a 0.25 μm diameter tungsten contact is probed with a narrow 20 kV beam, the W MNN signal intensity is determined to be only 70% of that emitted from a large tungsten structure. Target signal reduction coincides with increased signal contributions from the surrounding oxide.