Abstract

Six arsenic compounds including arsenocholine, arsenobetaine, dimethylarsinic acid, methylarsonic acid, arsenous acid and arsenic acid were separated by high-performance liquid chromatography (HPLC) on a Hamilton PRP-X100 anion-exchange column using isocratic elution and detected by inductively coupled plasma mass spectrometry (ICP-MS). This analytical procedure was applied to the speciation of arsenic compounds in human urine. The influence of urine matrix on the separation of arsenic compounds was evaluated and the determination of arsenic compounds was not hampered by the ArCl interference which has often been encountered in ICP-MS. Three human urine reference materials, SRM 2670 normal level, SRM 2670 elevated level and Lyphocheck urine metal control 1, were analyzed with respect to arsenic compounds by HPLC-ICP-MS. The results were found to be in good agreement with the certified total arsenic concentration in the reference materials. Six arsenic compounds were detected. Arsenobetaine was found to be present in all of the investigated human urine reference materials.