Abstract

A total of 20 heparinized pig kidneys were investigated at 10 intrapelvic hydrostatic pressure steps (0-90 mmHg). SWE (ARFI; Virtual TouchTM IQ, Siemens) measurements were taken at three different measuring regions and in two measuring sequences using a linear ultrasonography probe (9L4, Siemens). Median values of 10 shear-wave speed (SWS) measurements were calculated for each pressure step. Logarithmic transformed median SWS values were analyzed in a linear mixed model. SWS increased significantly with increasing intrapelvic pressure. Median SWS for all kidneys in both measuring sequences and all measuring regions was 1.47 m/s (interquartile range [IQR], 0.38 m/s) at 0 mmHg, 1.94 m/s (IQR, 0.42 m/s) at 30 mmHg, 2.07 m/s (IQR, 0.43 m/s) at 60 mmHg, 2.24 m/s (IQR, 0.49 m/s) at 90 mmHg. The correlation between pelvic pressure increase and median SWS values for the central parenchyma was significantly higher compared with the peripheral parenchyma. Acutely increased renal pelvic pressure correlates with increasing SWS values in ARFI elastography in an ex vivo porcine kidney model.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.