Abstract
Objective: To determine the reliability and agreement of different optical devices in terms of keratometry and pachymetry measurements in cross-linked (CXL) keratoconic eyes. Material and Methods: Thirty-eight CXL-treated keratoconic eyes were evaluated. Three repeated measurements were performed with Topcon KR-1 autorefractokeratometry, Sirius topography, Nidek AL-scan optical biometry and RTVue anterior segment optical coherence tomography (AS-OCT). The devices were compared in terms of pachymetry and keratometric values. Limits of agreement (LoA) between the devices were detected using Bland-Altman analysis. Intra-examiner reliability was obtained using the intraclass correlation coefficient (ICC). Results: While the mean central corneal thickness (CCT) and minimum corneal thickness (MCT) measured in AS-OCT were 456.37±41.52 μm and 431.89±43.37 μm, respectively, these values were 432.40±48.97 μm and 415.67±47.69 μm in topography (p0.97). When devices were compared in terms of the keratometric values and corneal astigmatism measurements, there was a significant difference among devices except between autorefractokeratometry and optical biometry (p0.99). Conclusion: Pachymetry and keratometry can be performed independently in each device with excellent reliability in CXL-treated keratoconic eyes. However, we consider that measurements made with different devices cannot be used interchangeably in these eyes.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.