Abstract
Appropriate or inappropriate ICD shock; what is the post-shock rhythm?
Highlights
These observations suggest that the sensed arrhythmia in the ventricular channel is not ventricular fibrillation, but sensing of atrial fibrillation
This observation can only be explained by a displacement of the ventricular lead towards the right atrium
The post-shock right atrial electrogram showed the simultaneous termination of atrial fibrillation, which suggests ventricular pacing
Summary
The post-shock right atrial electrogram showed the simultaneous termination of atrial fibrillation, which suggests ventricular pacing. Appropriate or inappropriate ICD shock; what is the post-shock rhythm? B. M. van Gelder1 · B. ter Burg1 · F.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.