Abstract

We explore the potential performance of soft x-ray spectrometers based on the use of varied-line-spacing spherical diffraction gratings (VLS-SG). The quantitative assessment is based on an optimization procedure to obtain both negligible optical aberrations at full illumination of the grating and a quasi linear focal curve. It involves high-order optical aberration cancellation to calculate the focal curves. We also examine the validity of small divergence closed-form formulas describing the light path function. Optimizing the optical and geometric parameters gives an ultimate resolving power, at 930 eV, of between 10,800 for a 3 m long instrument and 34,000 for an 11 m spectrometer according to the Rayleigh criterion. Typical fabrication tolerances would scale these values down by about 10%. The findings are validated by ray-tracing simulations.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.