Abstract
To evaluate the deposited dose distribution in a target, we have proposed to utilize the annihilation gamma-rays emitted from the positron emitters distributed in the target irradiated with stable heavy-ion beams. Verification of the one dimensional (1-D) dose distributions along and perpendicular to a beam axis was achieved through our previous works. The purpose of this work is to verify 3-D dose distributions. As the first attempt uniform PMMA targets were irradiated in simple rectangular parallelepiped shapes, and the annihilation gamma-rays were detected with a PET scanner. By comparing the detected annihilation gamma-ray distributions with the calculated ones the dose distributions were estimated. As a result the estimated positions of the distal edges of the dose distributions were in agreement with the measured ones within 1 mm. However, the estimated positions of the proximal edges were different from the measured ones by 5–9 mm depending on the thickness of the irradiation filed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.