Abstract
The ISO12103-1 A2 test dust was used as the reference in the calibration device of the atmospheric particle monitor to investigate the performance of the self-made standard dust in this study. The result indicated that when the aerosol concentration of the calibration device was 500μg/m3,400μg/m3, 200μg/m3,100μg/m3 and 50μg/m3, the stability of self-made standard dust was close to the ISO12103-1 A2 test dust. In addition, when the aerosol with high concentration of 500μg/m3, medium concentration of 200μg/m3 and low concentration of 50μg/m3, the homogeneity and obtained concentration of the self-made standard dust and the ISO12103-1 A2 test dust evaluated by filter membrane sampling were not much differencd. Therefore, the self-made standard dust can replace ISO12103-1 A2 test dust to be used in the calibration device of particle monitor.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: IOP Conference Series: Materials Science and Engineering
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.