Abstract

This study aims to develop a trigger operator based on the Overlap Volume Histogram (OVH) and examined its effectiveness in enhancing plan quality to minimize radiation-induced lung injury in postoperative radiotherapy for breast cancer. This trigger operator was applied for plan re-optimization to the previous Volumetric Modulated Arc Therapy (VMAT) plans of 16 left breast conserving surgery cases. These cases were categorized into a Contiguous Group (CG) and a Separated Group (SG) based on the relative position between the target and the Left-Lung (L-Lung). We investigated the changes in Vx, mean dose, and Normal Tissue Complication Probability (NTCP) values of organs-at-risk (OARs) before and after using the trigger operator. The Pairwise Sample T test was employed to evaluate the differences in indices between the two groups before and after optimizations. The trigger operator effectively initiated plan re-optimization. The values of V5, V10, V20, V30, and V40 of the L-Lung, as well as the mean dose of the heart, all decreased after re-optimization. The Pairwise Sample T test results showed statistically significant differences in the V20, V30, and V40 of the L-Lung in the CG (P < 0.01), and in the V5, V10, V20, V30, and V40 of the L-Lung in the SG (P < 0.01). Our findings suggest that the proposed trigger operator can improve plan quality, thereby reducing radiation-induced lung injury in postoperative radiotherapy for breast cancer.