Abstract

In order to develop a reliable, simple and routine method using small sample volume to determine norepinephrine (NE) and epinephrine (E) concentrations in plasma of humans and rodents, we utilize the ultrafiltration (UF) method by Ultrafree-MC filter device and a high-performance liquid chromatography equipped with electrochemical detector (HPLC-ECD) to detect NE and E. Optimum UF and HPLC conditions were as follows: the filter nominal molecular weight limit size is 30,000, the pH of added phosphate buffer to each plasma sample for UF is 3.0, and the mobile phase is 0.1 M phosphate buffer (pH 3)/acetonitrile (98:2) containing 0.05% sodium disulfite and 0.001% EDTA 2Na. The plasma samples and 1.0 M phosphate buffer (pH 3) containing 3,4-dihydroxybenzylamine (DHBA), as an internal standard, was mixed and poured into the UF units. After the centrifugation for 60 min at 13,000× g at 4 °C, the filtrate was directly injected into HPLC. The calibration curve of NE and E was linear for the concentrations studied (20–400 pg) with a correlation coefficient of >0.999. Intra-assay coefficients of variation for NE and E using this method were less than 3%. The method also correlated well with the well-established alumina method ( r=0.954). The present findings suggest that a newly-developed UF method with HPLC-ECD would apply successfully to measure plasma NE and E concentrations in humans and rodents.