Abstract

tris-[(4,7-diphenyl-1,10-phenanthroline)ruthenium(II)] dichloride (Ru(DPP)3Cl2), a fluorescent sensor which is sensitive to the amount of oxygen in the sample, was applied using the fluorescent optical respirometry (FOR) technique. The oxygen in the samples quenches the fluorescence. The fluorescence intensity depends on the metabolic rate of the viable microorganisms. The effect of DMSO and plant extracts on bacteria was determined by FOR. It was shown that the MIC values obtained by FOR were consistent with the results of the MIC determinations using the method of serial dilutions; at the same time, the effects of concentrations lower than the growth-inhibitory concentrations on microbial cells were demonstrated. The FOR method enables the detection of multiplying bacteria in sterile and non-sterile pharmaceutical preparations in real time, which significantly shortens the time required to obtain results and allows the introduction of repair processes in the production. This method also allows for quick, unambiguous detection and the counting of the viable cells of aerobic microorganisms in non-sterile pharmaceuticals.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call