Abstract
A quantitative, high throughput, fully automated diagnostic method for the detection of neutralizing anti-SARS-CoV-2 antibodies was developed on the Phadia system based on the interaction of SARS-CoV-2 S1 protein and the human ACE-2 receptor. This method was compared to the current state of the art plaque reduction neutralization test (PRNT) and a high correlation between the two methods was observed. Using a large cohort of blood samples from convalescent patients and controls the method displays very high sensitivity and specificity (99,8% and 99.99%, respectively). Neutralizing antibody titers of mRNA-1273 and BNT162b2-vaccinated persons can also be quantified with this method as well. This fully automated method provides the possibility to determine anti-SARS-CoV-2 neutralizing antibody concentrations in just 2 h.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.