Abstract
AbstractA new tin complex namely tetracyclohexyl tin(IV) (TCHT) was synthesized and used as the ion carrier for the construction of a highly selective salicylate sensor. This sensor shows a Nernstian response to salicylate ions over a very wide concentration (1.0 × 10−7–1.0 × 10−1 M) in a pH range of 5.5–10.5. The optimum selectivity and response could be obtained for a membrane incorporating 30% PVC, 61% BA, 3% of cationic additive (HTAB) and 6% of TCHT. The response time of the electrode is very short in the whole concentration range (15 s). The electrode also shows an excellent discriminating ability for salicylate ions with respect to the most common organic and inorganic anions including chloride, sulfate, nitrate, nitrite, cyanide, sulfite, iodide, thiocyanate, phosphate, acetate, oxalate, citrate, and tartarate ions. The detection limit of the proposed sensor is 8.0 × 10−8 M. The electrode was successfully used for determining the concentration of salicylate ion in synthetic serums.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.