Abstract

Bowl-shaped confinement was designed to increase signal stability for gas analysis. With normalization over the whole spectral area to alleviate the absorbed energy difference, our pulse-to-pulse RSDs for nitrogen and oxygen can reach 2.97% and 3.94% respectively, making LIBS analysis of gaseous samples more reliable.