Abstract

Aqueous solutions of the organic dye Rhodamine B were found to be useful in measuring radiation doses in the range 0.1–2kGy. Either spectrophotometric or spectrofluorometric measurements can be used. The dosimeter readout is unchanged for at least 60 days after irradiation if the solution is kept in the dark at room temperature. Effects of pH and the dye concentration on the dosimeter response were investigated. Radiation chemical yield and fluorescence quantum yield were also calculated.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.