Abstract

Ensuring quality of laboratory services is the need of the hour in the field of health care. Keeping in mind the revolution ushered by six sigma concept in corporate world, health care sector may reap the benefits of the same. Six sigma provides a general methodology to describe performance on sigma scale. We aimed to gauge our laboratory performance by sigma metrics. Internal quality control (QC) data was analyzed retrospectively over a period of 6 months from July 2009 to December 2009. Laboratory mean, standard deviation and coefficient of variation were calculated for all the parameters. Sigma was calculated for both the levels of internal QC. Satisfactory sigma values (>6) were elicited for creatinine, triglycerides, SGOT, CPK-Total and Amylase. Blood urea performed poorly on the sigma scale with sigma<3. The findings of our exercise emphasize the need for detailed evaluation and adoption of ameliorative measures in order to effectuate six sigma standards for all the analytical processes.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.