Abstract

Homovanillic acid (HVA) and vanillylmandelic acid (VMA) were selectively determined by quartz crystal nanobalance sensor in conjunction with net analyte signal (NAS)-based method called HLA/GO. An orthogonal design was applied for the formation of calibration and prediction sets including HVA, VMA, and some common and structurally similar urine compounds. The selection of the optimal time range involved the calculation of the NAS regression plot in any considered time window for each test sample. The searching of a region with maximum linearity of NAS regression plot (minimum error indicator) and minimum of predicted error sum of squares value was carried out by applying a moving window strategy. Based on the obtained results, the differences on the adsorption profiles in the time range between 1 and 300 s were used to determine mixtures of compounds by HLA/GO method. Several figures of merit like selectivity, sensitivity, analytical sensitivity, and limit of detection were calculated for both compounds. The results showed that the method was successfully applied for the determination of VMA and HVA.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.