Abstract

Ideally, medical x-ray imaging systems should be designed to deliver maximum image quality at an acceptable radiation risk to the patient. Quality assurance procedures are employed to ensure that these standards are maintained. A quality control protocol for direct digital radiography (DDR) systems is described and discussed. Software to automatically process and analyze the required images was developed. In this paper, the initial results obtained on equipment of different DDR manufacturers were reported. The protocol was developed to highlight even small discrepancies in standard operating performance.