Abstract

The current work delineates the development of a novel, rugged and sensitive stability-indicating risk-based HPLC method based on an analytical quality-by-design (QbD) approach for the concurrent estimation of naringin and pregabalin in dual-drug-loaded nanopharmaceuticals. Preliminary screening trials were conducted, along with systemic risk analysis, in order to identify the critical method attributes, namely injection volume, pH and acetonitrile content, that influence critical quality attributes. The Box-Behnken design was used to optimize the tailing factor as a response to pregabalin and naringin in a short run time. The chromatographic conditions were improved by running 17 experimental runs generated by design expert software. After analysing the optimized zone within the confines of the design space, the following chromatographic conditions were chosen: mobile phase water-acetonitrile adjusted to pH6.9 with phosphate buffer (80:20, %v/v), at flow rate of 1.0ml/min using a C18 analytical column at an isobestic wavelength of 212 nm. Furthermore, the optimized method was validated in accordance with International Conference on Harmonization guidelines and was found to be within the prescribed limits. The developed RP-HPLC method has a high degree of practical utility in in vivo and in vitro studies for the synchronous detection of pregabalin and naringin in pharmaceutical nanodosage forms such as protein-based nanoparticles, nanocrystals, polymeric nanoparticles and metallic nanoparticles.