Abstract

Chromatographic detection responses are recorded digitally. A peak is represented ideally by a Guassian distribution. Raising a Guassian distribution to the power ‘ n’ increases the height of the peak to that power, but decreases the standard deviation by √ n. Hence there is an increasing disparity in detection responses as the signal moves from low level noise, with a corresponding decrease in peak width. This increases the S/N ratio and increases peak to peak resolution. The ramifications of these factors are that poor resolution in complex chromatographic data can be improved, and low signal responses embedded at near noise levels can be enhanced. The application of this data treatment process is potentially very useful in 2D-HPLC where sample dilution occurs between dimension, reducing signal response, and in the application of post-reaction detection methods, where band broadening is increased by virtue of reaction coils. In this work power functions applied to chromatographic data are discussed in the context of (a) complex separation problems, (b) 2D-HPLC separations, and (c) post-column reaction detectors.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.