Abstract
Figures of merit (limit of detection, sensitivity and precision) for eight elements (Ag, Cd, Pb, Mn, Sn, Tl, As and Se) were obtained using furnace atomization plasma emission spectrometry (FAPES) in combination with atomization of samples from a L'vov platform and the use of a Pd modifier (0.5 µg). A 50 W He plasma was utilized. Precision of replicate integrated signals averaged 2.8%(range 1.7–4.3%) for injection, by hand, of samples at concentrations 13–70-fold above the detection limit. Estimated limits of detection and sensitivities for integrated intensities improved 3–10- and 6–17-fold, respectively, for volatile elements (Cd, Pb and Ag) compared with sample atomization from the tube wall. Added NaCl, up to 24 µg, had no effect on the recovery of the integrated signal from Pb when a 100 W plasma was used.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.