Abstract
Two ion-selective sensors of the plastic membrane type were prepared for the determination of oxybutynin hydrochloride (OxCl). They depend on the incorporation of the ion-associates with phosphotungestic acid or phosphomolybdic acid in a PVC matrix. A comparative study is made between their performance characteristics in batch and FIA conditions. The sensors have nearly the same usable concentration, temperature and pH range. They have a wide range of selectivity and can be applied for the determination of the relevant drug with nearly the same precision and accuracy in vitro. Dissolution testing was applied using the sensors; this offers a simple, rapid, cheap way out of sophisticated and high cost instruments used in the pharmacopeial method using HPLC. The investigated drug was determined in its pure and pharmaceutical preparations. The results were accurate and precise, as indicated by the recovery values and coefficients of variation.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.