An ultrasensitive sandwich-type electrochemiluminescence (ECL) immunosensor was designed for carcinoembryonic antigen (CEA) detection. This immunosensor was developed using nanoporous Pd as excellent labels and Ag/graphene (Ag–G) nanocomposite as biosensing substrates. Ag–G nanocomposite was prepared by simultaneous reduction of graphene oxide and Ag ions with a rapid and efficient one-step approach and served as an effective matrix for primary antibodies attachment. The nanoporous Pd obtained from dealloying Pd/Al alloy is an excellent catalyst for the oxidation of H2O2 to generate O2 due to its large specific surface area and high catalytic activity. O2 could improve the ECL performance of peroxydisulfate through reducing the strong oxidant SO4•−. The proposed immunosensor successfully fulfilled the ultrasensitive detection of CEA in the range from 1pgmL−1 to 500ngmL−1 with a detection limit of 0.6pgmL−1. Finally, our ECL immunosensor has advantages of high sensitivity, specificity and stability, and could be used to monitor CEA level in human serum with satisfactory results.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.