Abstract
AbstractAn amperometric detector based on the chemical modification of Nafion and cobalt (II) hexacyanoferrate (II, III) thin film (Nafion/Co‐CN‐Fe) onto a glassy carbon (GC) electrode was firstly developed for the determination of electroinactive cations (Li+, Na+, K+, Rb+, Cs+ and NH4+) in single column ion chromatography. A set of well‐defined peaks of electroinactive cation was obtained. The relative standard deviations (RSDs) of chromatographic peak height (nA) for these cations were all below 3.8%. The cations were detected conveniently in the linear concentration range of 6.0 × 10−6‐5.0 × 10−3 mol/L and their correlation coefficients were all above 0.99. The detection limits of the cations were 9.2 × 10−6 mol/L for Li+, 3.4 × 10−6 mol/L for Na+, 6.3 × 10−7 mol/L for K+, 7.8 × 10−7 mol/L for Rb+, 6.2 × 10−7 mol/L for Cs+ and 6.2 × 10−6 mol/L for NH4+, at a signal‐noise ratio of 3. The method was quick, sensitive, simple and was successfully applied to the analysis of rainwater samples. The electrode was stable for a 2 week period of operation with no evidence of chemical or mechanical deterioration.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.