Abstract
The article describes the use of multi-criteria optimization methods during the calibration of digital multimeters. Currently, calibration is based on a single measurement of a specific value. The aim of this research was to confirm the possibility of using a series of measurements in order to reduce the measurement uncertainty without significantly extending the calibration time. The automatic measurement loading laboratory stand used during the conducted experiments was essential to obtain results that allowed confirming the thesis. This article presents the applied optimization methods and the results of the calibration of sample digital multimeters obtained thanks to them. As a result of the research, it was found that the use of a series of measurements increased the accuracy of the calibration, reduced the measurement uncertainty, and shortened the calibration time compared to traditional methods.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.